Accuracy of Various Molecular Weight Determination Methods

Method of Determination Standard Error of Estimate
Mw by LALLS (Mw above 50,000) +/- 5 to 10 %
Mw by LALLS (Mw below 50,000) +/- 10 to 15 %
Mn by LALLS +/- 20 to 30 %
Mw by MALLS (Mw above 50,000) +/- 3 to 10 %
Mw by MALLS (Mw below 50,000) +/- 10 to 15 %
Mn by MALLS +/- 20 to 30 %
Mn by Membrane Osmometry (Mw above 50,000) +/- 5 to 10 %
Mn by Vapor Pressure Osmometry (Mw 15,000 - 30,000) +/- 10 to 15 %
Mn by Vapor Pressure Osmometry (Mw below 15,000 ) +/- 2 to 5 %
Mw by GPC/SEC +/- 3 to 10 %
Mn by GPC/SEC +/- 10 to 15 %
Mz by GPC/SEC +/- 10 to 15 %
Mp by GPC/SEC +/- 3 to 10 %
Mw by GPC/DV Universal Calibration +/- 5 to 10 %
Mn by GPC/DV Universal Calibration +/- 10 to 15 %
Mz by GPC/DV Universal Calibration +/- 10 to 15 %
Mp by GPC/DV Universal Calibration +/- 5 to 10 %
Mn by End Group Analysis (Mw below 10,0000) +/- 10 to 15 %
Mz by Ultra Centrifuge (Mw above 10,000) +/- 10 to 15 %
Mv by Viscotek Viscometer Y501-C +/- 5 to 10 %
Intrinstic Viscosity by Viscometer Y501-C +/- 5 to 10 %